Posniak H V, Olson M C, Demos T C, Pierce K L, Kalbhen C L
Department of Radiology, Loyola University Medical Center, Maywood, IL 60153.
AJR Am J Roentgenol. 1994 Nov;163(5):1073-7. doi: 10.2214/ajr.163.5.7976877.
Chest and abdominal CT scans using 1.0-sec scan times are often limited by motion in patients who are unable to hold their breath. With our scanner we can obtain images in 0.6 sec (partial scan) that use data from 225 degrees instead of the 360 degrees used for 1.0-sec scans. The purpose of this study was to assess whether the quality of images of the chest and abdomen in patients on mechanical pulmonary ventilators who could not breath-hold could be improved if images were taken using a scan time of 0.6 sec rather than 1.0 sec.
Thirty patients who were being treated with mechanical pulmonary ventilation with indications for chest or abdominal CT or both were scanned with a scan time of 1.0 sec. At the end of the examination, additional 0.6-sec images were taken at three or four levels. Images obtained with these two scan times were filmed at similar levels and windows, and anatomically matched levels were compared. The images were masked and independently evaluated by three radiologists for motion, noise, artifact, and overall image quality. Each parameter was rated using a scale of 1 to 4. Using Kendall's tau correlation, there was no significant difference between the radiologists in the grading of individual parameters or of overall image quality. Therefore, the average of the scores of the three radiologists was used. Statistical analysis was done using repeated measures multivariate analysis of variance.
Images obtained in 0.6 sec had significantly less motion (p < .001) but more noise (p < .001) than those obtained in 1.0 sec. We found no statistically significant difference in artifacts between the two scan times. Overall image quality was judged to be significantly better on scans obtained in 0.6 sec than on scans obtained at 1.0 sec (p < .001), in spite of the greater noise on 0.6-sec images. The higher quality of 0.6-sec images was most noticeable for lung window settings.
Our results indicate that scans taken in 0.6 sec (partial scans) provide better quality images than those obtained in 1.0 sec in patients being treated with mechanical pulmonary ventilation who cannot breath-hold. This technique may be useful not only in this population but in all patients who are unable to cooperate with breath-holding instructions.
对于无法屏气的患者,采用1.0秒扫描时间进行胸部和腹部CT扫描常因运动而受限。使用我们的扫描仪,我们可以在0.6秒(局部扫描)内获得图像,该扫描使用来自225度的数据,而非1.0秒扫描所使用的360度数据。本研究的目的是评估对于使用机械通气且无法屏气的患者,采用0.6秒而非1.0秒的扫描时间获取图像,是否能改善胸部和腹部图像的质量。
30例因胸部或腹部CT或两者检查而接受机械通气治疗的患者,先以1.0秒的扫描时间进行扫描。检查结束时,在三到四个层面额外采集0.6秒的图像。这两种扫描时间获得的图像在相似层面和窗宽下进行拍摄,并对解剖学匹配层面进行比较。图像进行遮蔽处理后,由三位放射科医生独立评估其运动、噪声、伪影及整体图像质量。每个参数采用1至4分的评分标准。使用肯德尔tau相关性分析发现,放射科医生在各个参数及整体图像质量评分上无显著差异。因此,采用三位放射科医生评分的平均值。采用重复测量方差分析进行统计分析。
与1.0秒获得的图像相比,0.6秒获得的图像运动伪影明显更少(p < .001),但噪声更多(p < .001)。我们发现两种扫描时间的伪影在统计学上无显著差异。尽管0.6秒图像噪声更大,但总体图像质量在0.6秒扫描时被判定显著优于1.0秒扫描时(p < .001)。在肺窗设置下,0.6秒图像更高的质量最为明显。
我们的结果表明,对于使用机械通气且无法屏气的患者,0.6秒(局部扫描)获得的扫描图像质量优于1.0秒获得的图像。该技术不仅对这一群体有用,对所有无法配合屏气指令的患者也可能有用。