Moreau E, Philippé J, Kint J, Benoit Y, Leroux-Roels G
Laboratorium voor Klinische Biologie, Universitair Ziekenhuis, Gent, Belgium.
Eur J Clin Chem Clin Biochem. 1994 Jul;32(7):563-6.
The Coulter STKS and the Sysmex NE-1500 were evaluated for their performance in paediatric samples. Special attention was paid to the ability of both instruments to detect white blood cell morphology abnormalities. Therefore a population with a high frequency of pathologies was selected. The studied population included children aged between 1 month and 14 years, 45% of which suffered from oncological and/or haematological diseases. Two hundred and forty blood samples were analysed. The leukocyte differentiation results from the Coulter STKS and the Sysmex NE-1500 were compared with the manual microscopic method. The differentials compared very well for neutrophils, lymphocytes and eosinophils. Sensitivity and specificity for the detection of blasts, immature granulocytes, left shift and atypical lymphocytes were examined. While specific morphology flagging showed a rather low sensitivity, blasts and immature granulocytes were mostly detected by the two systems, taking into account white blood cell alarm flags other than the two mentioned above. The overall sensitivity for the detection of white blood cell abnormalities was better with the Coulter STKS, whereas the Sysmex NE-1500 showed a higher specificity. In an ancillary study which was performed exclusively on neonatal samples, of which 80% originated from preterm infants, we found a limited sensitivity for the detection of low percentages of nucleated red blood cells for both analysers. The high frequency of abnormalities, together with the limited sensitivity mentioned above, prompted us to review all neonatal samples in the future.
对库尔特STKS血细胞分析仪和希森美康NE - 1500血细胞分析仪在儿科样本中的性能进行了评估。特别关注了这两种仪器检测白细胞形态异常的能力。因此,选择了一个疾病发生率高的人群。研究人群包括1个月至14岁的儿童,其中45%患有肿瘤和/或血液疾病。分析了240份血样。将库尔特STKS血细胞分析仪和希森美康NE - 1500血细胞分析仪的白细胞分类结果与手工显微镜检查方法进行了比较。中性粒细胞、淋巴细胞和嗜酸性粒细胞的分类结果比较吻合。对原始细胞、未成熟粒细胞、核左移和非典型淋巴细胞的检测敏感性和特异性进行了检查。虽然特定形态标记的敏感性相当低,但考虑到除上述两种之外的白细胞报警标记,原始细胞和未成熟粒细胞大多能被这两种系统检测到。库尔特STKS血细胞分析仪检测白细胞异常的总体敏感性更好,而希森美康NE - 1500血细胞分析仪显示出更高的特异性。在一项专门针对新生儿样本的辅助研究中,其中80%来自早产儿,我们发现这两种分析仪检测低比例有核红细胞的敏感性有限。异常的高发生率以及上述有限的敏感性促使我们在未来对所有新生儿样本进行复查。