血液学中三种自动血细胞计数器的比较及同步评估:库尔特STKS、希森美康NE 8000、Technicon H-2

[Comparative and simultaneous evaluation of three automated counters in hematology: Coulter STKS, Sysmex NE 8000, Technicon H-2].

作者信息

Cart-Lamy P, Pernod G, Pirraud M, Drevait J M, Caillat D, Polack B, Kolodié L

机构信息

Groupe Bioxel, Meylan, France.

出版信息

Ann Biol Clin (Paris). 1993;51(6):627-35.

PMID:8172402
Abstract

We evaluated and compared three automated blood cell counters, Coulter STKS, Sysmex NE 8000 and Technicon H-2. These perform both a complete blood cell count and a full white cell differential count. Carry-over was found to be acceptable, except for the leucocyte count by the Sysmex NE 8000 (1.5%; p < 0.001). Linearity over a wide concentration range for all of the measured parameters, haemoglobin, red blood cells, white blood cells and platelets, was excellent (r > 0.98). Within-run precision was verified with 22 samples covering a wide range of cell counts by repeated analysis (n = 20) of the same sample. Coefficients of variation (CV) were acceptable, < 4% for values within the normal range. The CVs for eosinophils and basophils were less good but without clinical impact. Analysis of 270 samples showed that the Coulter counter indicated the presence of atypical cell populations more frequently (19%; p = 0.02). Message of qualitative abnormalities displayed by the three analysers were often discordant. These three blood cell counters differed in their sensitivity and positive predictive value for detecting abnormal blood cells rather than in their specificity and negative predictive value. Other operating aspects of three instruments that are important in haematological practice are documented.

摘要

我们评估并比较了三种自动血细胞计数器,即库尔特STKS、希森美康NE 8000和Technicon H-2。这些仪器均可进行全血细胞计数和完整的白细胞分类计数。除希森美康NE 8000的白细胞计数外(1.5%;p<0.001),交叉污染情况可接受。对于所有测量参数,即血红蛋白、红细胞、白细胞和血小板,在很宽的浓度范围内线性关系良好(r>0.98)。通过对同一22个涵盖广泛细胞计数范围的样本进行重复分析(n=20)来验证批内精密度。变异系数(CV)可接受,正常范围内的值<4%。嗜酸性粒细胞和嗜碱性粒细胞的CV稍差,但无临床影响。对270个样本的分析表明,库尔特计数器更频繁地显示存在非典型细胞群(19%;p=0.02)。这三种分析仪显示的定性异常信息往往不一致。这三种血细胞计数器在检测异常血细胞方面的敏感性和阳性预测值不同,而不是特异性和阴性预测值。记录了这三种仪器在血液学实践中重要的其他操作方面。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验