A particle sizing counter, elaborated by the Institute of Cytology and Scientific-Industry Union "Analytpribor", was evaluated for its effectiveness in analysing different cell suspensions. The mean error of cell concentration measurement is about 3% within the 1-15 x 10(4) ml-1 interval. Cells can be found in suspension at a concentration of 2 x 10(2) ml-1. Cell size can be measured reproducibly within 10% for the interval 3-150 microns. The absolute value of cell size can be obtained after the device graduation with independently sized cells. Such a graduation differs by 5-10% from the graduation with latexes.
由细胞学研究所和“Analytpribor”科学工业联盟研制的一种颗粒大小计数器,针对其分析不同细胞悬液的有效性进行了评估。在1 - 15×10⁴ ml⁻¹区间内,细胞浓度测量的平均误差约为3%。在浓度为2×10² ml⁻¹的悬液中可检测到细胞。在3 - 150微米区间内,细胞大小的测量可重复性在10%以内。使用独立确定大小的细胞对该装置进行校准后,可获得细胞大小的绝对值。这种校准与使用乳胶进行的校准相差5 - 10%。