Relation between the size of lesions and arrhythmias produced by microwave catheter ablation with a special electrode device.

作者信息

Ikeda T, Sugi K, Enjoji Y, Kasao M, Abe R, Ninomiya K, Yabuki S, Yamaguchi T

机构信息

Third Department of Internal Medicine, Ohashi Hospital, Toho University School of Medicine, Tokyo, Japan.

出版信息

Jpn Circ J. 1994 Mar;58(3):214-21. doi: 10.1253/jcj.58.214.

Abstract

The purpose of this study was to examine the effects of microwave catheter ablation of ventricular myocardium. Microwave energy with a frequency of 2450 MHz was delivered via a coaxial catheter with an electrode ball tip. Microwave energy was applied to canine isolated left ventricular endocardium in vitro and to 6 anesthetized dogs in vivo at 50 watts for 15-150 sec. Ventricular arrhythmia was not observed during ablation in any of the dogs when microwave energy was applied for less than 45 sec. When the duration of microwave ablation was greater than 45 sec, ventricular premature contractions were observed in all of the dogs. Nonsustained ventricular tachycardia developed when the duration of microwave delivery was greater than 90 sec. After the cessation of ablation, ventricular arrhythmias did not occur and ventricular programmed stimulation did not induce ventricular tachycardia in any of the dogs. Except for ventricular arrhythmia, no declines in the hemodynamic status were observed in any of the 6 dogs. The size of the ablated lesion was significantly greater as the duration of ablation was increased (p < 0.05). When ablation lasted for more than 120 sec, the coagulation layer was extended to the epicardium in all 6 dogs. The results of this study suggest that microwave ablation is feasible for the treatment of tachyarrhythmias from deep focus of ventricular myocardium with relatively small proarrhythmic effects.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验