Moussavi M, Schwan H P, Sun H H
Institute of Biomedical Engineering & Science, Drexel University, Philadelphia, PA 19104.
Med Biol Eng Comput. 1994 Mar;32(2):121-5. doi: 10.1007/BF02518907.
Steady-state measurements of the electrical properties of electrodes immersed in physiological saline were taken with small- and large-amplitude sinusoidal voltages (30 m V-1 V) in the frequency range 0.2 mHz-2 Hz. Impedance magnitude and phase, and the first four Fourier series coefficients of the polarised current were measured. The dependence of the polarisation admittance on input current intensity was modelled with the linear relationship proposed by Schwan. This model predicts harmonics in the non-linear range for AC overpotentials up to several hundred millivolts. Observed values deviated from theoretical values. Reasons for such deviations are discussed. The study of harmonic distortion appears to be a useful tool to monitor and predict the non-linear behaviour of the interface and other dielectric phenomena.
使用小幅度和大幅度正弦电压(30 mV - 1 V)在0.2 mHz - 2 Hz频率范围内对浸入生理盐水中的电极进行电性能的稳态测量。测量了阻抗幅值和相位以及极化电流的前四个傅里叶级数系数。用施万提出的线性关系对极化导纳对输入电流强度的依赖性进行建模。该模型预测了高达数百毫伏的交流过电位在非线性范围内的谐波。观测值与理论值存在偏差。讨论了这种偏差的原因。谐波失真研究似乎是监测和预测界面非线性行为及其他介电现象的有用工具。