A clinical study of an electronic constant force periodontal probe.

作者信息

Tupta-Veselicky L, Famili P, Ceravolo F J, Zullo T

机构信息

Department of Periodontics, West Virginia University, Morgantown.

出版信息

J Periodontol. 1994 Jun;65(6):616-22. doi: 10.1902/jop.1994.65.6.616.

Abstract

Automated probing systems have been developed to provide a more precise method of evaluating periodontal pocket depths and attachment levels. The purpose of this study was to assess the clinical performance (reproducibility, time, and comfort) of a new electronic probe (E.P.) compared to the UNC 15 conventional probe (C.P.). Paired measurements 2 hours apart were performed by one examiner on 20 patients with moderate to advanced localized or generalized adult periodontitis. Both the E.P. and C.P. were used on each patient in a random manner. Overall reproducibility (+/- 1.5 mm) was: E.P. 94% (n = 1181); C.P. 96.5% (n = 1254); E.P./C.P. used interchangeably 82.4% (n = 830). In assessing the reproducibility of bleeding on probing, using the Wilcoxon matched pairs signed ranks test, only the mid-facial surface, when using the C.P. exhibited differences between measurements (P < 0.017). Paired t test revealed E.P. took significantly longer per exam (4 minutes, 46 seconds). Comfort levels, evaluated with a visual analog scale, were not significantly different between probes as shown by the Mann Whitney U test. The data suggest that, in general, there was no major significant difference in reproducibility measurements between the E.P. and the C.P. The E.P. took more time. Comfort levels were similar for both probes.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验