Krohg-Sørensen K, Line P D, Kvernebo K
Unit of Vascular Surgery, Aker Hospital, Oslo, Norway.
Eur J Surg. 1993 May;159(5):293-9.
To evaluate laser Doppler flowmetry (LDF) and tissue oximetry (TpO2) for intraoperative assessment of colonic perfusion in pigs.
Eight pigs weighing from 19-22 kg.
Relationships of laser Doppler flux and TpO2 to mean arterial pressure (MAP) were measured in porcine sigmoid colon. The influence of different fractional inspired oxygen concentrations (FiO2) and the variability of the laser Doppler flux were evaluated.
Highly significant reproducible non-linear perfusion compared with pressure relations were found with both methods (r > or = 0.9, p < 0.0001). There was a rapid decrease when MAP fell below 90 mmHg. The dependency of TpO2 on FiO2 was omitted by using the ratio TpO2:PaO2. The coefficient of variation of repeated LDF recordings of 30 second duration with the hand held probe was 0.11 (0.07-0.16) (median and 95% confidence interval). Three recordings gave a reproducible mean flux with an estimated precision of 10% either way. Simultaneous recordings with sutured and hand-held probes correlated well (r > or = 0.9, p < or = 0.01).
Both methods yielded quantitative and reproducible intraoperative assessments of colonic perfusion, and are applicable for clinical use.
评估激光多普勒血流仪(LDF)和组织血氧饱和度仪(TpO2)在猪结肠灌注术中评估中的应用。
8头体重19 - 22千克的猪。
在猪乙状结肠中测量激光多普勒血流与TpO2与平均动脉压(MAP)的关系。评估不同吸入氧分数(FiO2)的影响以及激光多普勒血流的变异性。
两种方法均发现与压力关系相比具有高度显著的可重复非线性灌注(r≥0.9,p<0.0001)。当MAP降至90 mmHg以下时,出现快速下降。通过使用TpO2:PaO2比值消除了TpO2对FiO2的依赖性。手持探头进行的30秒持续时间的重复LDF记录的变异系数为0.11(0.07 - 0.16)(中位数和95%置信区间)。三次记录给出了可重复的平均血流,估计精度在任何一种情况下均为10%。缝合探头和手持探头的同步记录相关性良好(r≥0.9,p≤0.01)。
两种方法均能在术中对结肠灌注进行定量且可重复的评估,适用于临床应用。