Thaete F L, Fuhrman C R, Oliver J H, Britton C A, Campbell W L, Feist J H, Straub W H, Davis P L, Plunkett M B
Department of Radiology, University of Pittsburgh, PA 15261-0001.
AJR Am J Roentgenol. 1994 Mar;162(3):575-81. doi: 10.2214/ajr.162.3.8109499.
The purpose of this study was to compare observer performance in interpreting high-quality, digitally acquired computed radiographs of the chest displayed on either laser-printed radiographs or a workstation with observer performance in interpreting conventional chest radiographs.
We performed a receiver-operating-characteristic study in which the presence or absence of five abnormalities was determined by nine experienced radiologists on 310 posteroanterior radiographs of the chest displayed in three forms: conventional radiographs, laser-printed films of digital radiographs, and digital radiographs on a high-resolution workstation.
The results of our study suggest that observer performance with laser-printed films of digital radiographs obtained with high-resolution (4K x 5K), high-contrast sensitivity (12 bits) and appropriate exposure is comparable to observer performance with conventional radiographs. Observer performance with digital radiographs displayed on the workstation was found to be significantly lower for abnormalities that contained high-frequency and low-contrast information (e.g., interstitial disease and pneumothorax).
Computed radiography technology can produce image quality that is adequate for interpreting posteroanterior radiographs of the chest. Observer performance is not as good when radiographs displayed on workstations are used to diagnose specific abnormalities.
本研究旨在比较观察者对以激光打印胸片或工作站形式显示的高质量数字化采集胸部计算机X线片的解读表现,与对传统胸部X线片的解读表现。
我们进行了一项接受者操作特征研究,由9名经验丰富的放射科医生在310张胸部后前位X线片上判断5种异常情况的有无,这些X线片以三种形式显示:传统X线片、数字X线片的激光打印胶片以及高分辨率工作站上的数字X线片。
我们的研究结果表明,观察者对通过高分辨率(4K×5K)、高对比度敏感度(12位)以及适当曝光获得的数字X线片的激光打印胶片的解读表现,与对传统X线片的解读表现相当。对于包含高频和低对比度信息的异常情况(如间质性疾病和气胸),发现观察者对工作站上显示的数字X线片的解读表现明显较低。
计算机X线摄影技术能够产生足以用于解读胸部后前位X线片的图像质量。当使用工作站上显示的X线片诊断特定异常时,观察者的表现不佳。