两种评估人体体外循环期间脑静脉血氧血红蛋白饱和度系统的比较。

A comparison of two systems for assessing cerebral venous oxyhaemoglobin saturation during cardiopulmonary bypass in humans.

作者信息

Brown R, Wright G, Royston D

机构信息

Department of Anaesthesia, Harefield Hospital, Middlesex.

出版信息

Anaesthesia. 1993 Aug;48(8):697-700. doi: 10.1111/j.1365-2044.1993.tb07184.x.

Abstract

The Somanetics Invos 3,100 cerebral oximeter is a new noninvasive device which measures the regional oxygen saturation of haemoglobin within the cerebral microvasculature by infrared spectroscopy. It was used in nine patients undergoing elective cardiac surgery and its results were compared with measurements of jugular venous bulb saturations obtained by the Oximetrix Opticath oximetry system. The index value of jugular venous bulb saturation was obtained by analysis of jugular bulb blood in an Il282 cooximeter. The cerebral oximeter was less accurate and precise (standard deviation of difference 14.1%) than the Oximetrix system (standard deviation of the difference 2.65%) and also demonstrated a systematic error in bias unrelated to cerebral perfusion pressure. The cerebral oximeter may therefore be less useful than jugular venous bulb saturation in the clinical management of patients undergoing cardiopulmonary bypass.

摘要

索曼etics Invos 3100脑氧饱和度仪是一种新型无创设备,它通过红外光谱法测量脑微血管内血红蛋白的局部氧饱和度。该仪器用于9例择期心脏手术患者,并将其结果与通过血氧测定仪Opticath血氧测定系统获得的颈静脉球饱和度测量结果进行比较。颈静脉球饱和度的指标值是通过在Il282型血液气体分析仪中分析颈静脉球血液获得的。与血氧测定仪系统(差异标准差为2.65%)相比,脑氧饱和度仪的准确性和精密度较低(差异标准差为14.1%),并且在与脑灌注压无关的偏差方面也表现出系统误差。因此,在体外循环患者的临床管理中,脑氧饱和度仪可能不如颈静脉球饱和度有用。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验