Suppr超能文献

圣犹达医疗公司19毫米和21毫米血流动力学增强型双叶心脏瓣膜流体动力学功能的对比研究。

Comparative study of the hydrodynamic function of the size 19mm and 21mm St. Jude Medical Hemodynamic Plus Bileaflet Heart valves.

作者信息

Fisher J

机构信息

Department of Mechanical Engineering, University of Leeds, United Kingdom.

出版信息

J Heart Valve Dis. 1994 Jan;3(1):75-80.

PMID:8162222
Abstract

The hydrodynamic function of the St. Jude Medical Hemodynamic Plus Bileaflet Heart valve has been compared to the function of the standard St. Jude valve, the Carbomedics, Duromedics and the Monostrut valves in a pulsatile flow simulator. Size 19mm and 21mm tissue annulus diameter valves have been studied for use in small aortic roots. The size 19mm St. Jude hemodynamic Plus had lower forward flow pressure drops and lower total energy losses than the other valves studied. The size 19mm St. Jude Hemodynamic Plus valve should provide lower forward flow pressure drops and energy losses for the patient with a small aortic root, than the other types of prostheses included in the study.

摘要

在脉动流模拟器中,已将圣犹达医疗公司的血流动力学增强型双叶心脏瓣膜的流体动力学功能与标准圣犹达瓣膜、卡博梅迪克斯瓣膜、杜罗梅迪克斯瓣膜和单支柱瓣膜的功能进行了比较。已对尺寸为19毫米和21毫米组织瓣环直径的瓣膜进行了研究,以用于小主动脉根部。与所研究的其他瓣膜相比,尺寸为19毫米的圣犹达血流动力学增强型瓣膜具有更低的前向血流压降和更低的总能量损失。对于主动脉根部较小的患者,尺寸为19毫米的圣犹达血流动力学增强型瓣膜应比该研究中包含的其他类型假体提供更低的前向血流压降和能量损失。