Frame-mounted porcine valve bioprostheses: preparation during aortic-root dilation. Biomechanics and design considerations.

作者信息

Butterfield M, Fisher J, Lockie K J, Davies G A, Watterson K

机构信息

Department of Mechanical Engineering, University of Leeds, United Kingdom.

出版信息

J Thorac Cardiovasc Surg. 1993 Dec;106(6):1181-8.

PMID:8246557
Abstract

The leaflet geometry and function of frame-mounted porcine bioprostheses prepared with dilation of the aortic root during frame mounting was investigated. The diameter of the porcine aortic root increased by 47% when dilated with a pressure of 120 mm Hg. In the absence of pressure dilation, the ratio of circumferential leaflet length to the radius of the aortic root (S/RA) was 2.7 +/- 0.1, and the angle of inclination of the leaflet to the base of the valve was 17 +/- 4.3 degrees. In this condition, the leaflet geometry was similar to that of some second-generation porcine bioprostheses, which demonstrated high open-leaflet bending strains at the commissures. Dilation of the porcine root with 120 mm Hg reduced the value of S/RA to 1.84 and produced a triangular open-leaflet configuration with minimum open-leaflet bending strains. Open-leaflet bending strains were also reduced in two prototype frame-mounted valves prepared with partial dilation of the aortic root, which had an S/RA ratio of less than 2.3 and a leaflet angle greater than 27 degrees. The study indicates that the next generation of porcine bioprostheses should be prepared with at least 17% dilation of the aortic root during frame mounting and with zero pressure difference across the leaflets during fixation. This can be achieved by applying an equal hydrostatic pressure to both sides of the valve leaflets during fixation and frame mounting.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验