MacLean G K, Aiken P A, Adams W A, Mussivand T
Cardiovascular Devices Division, University of Ottawa Heart Institute, Ontario, Canada.
ASAIO J. 1993 Jul-Sep;39(3):M423-6.
The mismatching of battery cell capacities within a rechargeable battery pack can result in reduced cycle life and operating time, as well as an increased safety risk due to gas release, which is a concern when used in implantable medical devices, such as a ventricular assist device (VAD). Two groups of four nickel/cadmium (Ni/Cd) battery packs containing 10 cells each (i.e., 12 volt packs) were evaluated for their cycle life, safe operating time, and the impact of mismatched cell capacities within the pack. Four control battery packs containing capacity matched cells (within +/- 0.4% of mean) and four battery packs each containing one low capacity cell (4.7-10.7% below mean) together with nine high capacity matched cells (within +/- 0.6% of mean) were charge/discharge cycled at 37 degrees C using a 9.7 W average pulsatile discharge load and a 30 minute discharge time on each cycle. Periodically, the battery packs were fully discharged down to 10 volts and each cell was monitored for negative voltages. The capacity matched and mismatched packs yielded similar performance curves, with an initial decline in mean total operating time until a plateau was reached of 41.1 +/- 1.5 and 41.8 +/- 1.8 minutes of discharge time, respectively. The mean total operating times of the matched and mismatched packs began to decline again after approximately 670 and 635 cycles, respectively. Cell voltage monitoring showed no negative voltages, indicating that no gases were released from the cells.(ABSTRACT TRUNCATED AT 250 WORDS)
可充电电池组内电池单元容量不匹配会导致循环寿命和工作时间缩短,同时由于气体释放会增加安全风险,这在用于植入式医疗设备(如心室辅助装置,VAD)时是一个问题。对两组各含10个电池单元(即12伏电池组)的镍镉(Ni/Cd)电池组进行了循环寿命、安全工作时间以及电池组内电池单元容量不匹配影响的评估。四个包含容量匹配电池单元(在平均值的±0.4%范围内)的对照电池组和四个每个包含一个低容量电池单元(比平均值低4.7 - 10.7%)以及九个高容量匹配电池单元(在平均值的±0.6%范围内)的电池组,在37摄氏度下使用9.7瓦平均脉动放电负载且每个循环放电时间为30分钟进行充放电循环。定期将电池组完全放电至10伏,并监测每个电池单元的负电压。容量匹配和不匹配的电池组产生了相似的性能曲线,平均总工作时间最初下降,直到分别达到41.1±1.5分钟和41.8±1.8分钟的放电时间平台期。匹配和不匹配电池组的平均总工作时间分别在大约670次和635次循环后再次开始下降。电池单元电压监测未显示负电压,表明电池单元未释放气体。(摘要截选至250字)