de Vries P M, Langendijk J W, Kouw P M, Visser V, Schneider H
Department of Internal Medicine, Free University Hospital, Amsterdam, The Netherlands.
Med Biol Eng Comput. 1993 Sep;31(5):445-8. doi: 10.1007/BF02441978.
A study was designed to explore the possibility of detecting the haematocrit of blood by means of admittance measurements. The admittance and phase angle of blood kept in a measuring cell were determined at various frequencies between 60 kHz and 24 MHz. A reliable and accurate estimation of haematocrit was obtained in two ways. First, low-frequency admittance, high-frequency admittance and a factor x, which was the conductive percentage of cell content, were used. Secondly, the maximum phase angle was used. Both methods can be applied to obtain continuous on-line information about haematocrit for blood volume control during haemodialysis.
一项研究旨在探索通过导纳测量来检测血液血细胞比容的可能性。在60千赫至24兆赫的不同频率下,测定了置于测量池中血液的导纳和相位角。通过两种方法获得了对血细胞比容可靠且准确的估计。第一种方法,使用低频导纳、高频导纳以及一个表示细胞内容物导电百分比的因子x。第二种方法,使用最大相位角。这两种方法均可用于在血液透析期间获取有关血细胞比容的连续在线信息,以控制血容量。