Suppr超能文献

用于临床听力测定的听力测试室中的环境噪声水平。

Ambient noise levels in audiometric test rooms used for clinical audiometry.

作者信息

Frank T, Williams D L

机构信息

Department of Communication Disorders, Penn State University, University Park.

出版信息

Ear Hear. 1993 Dec;14(6):414-22. doi: 10.1097/00003446-199312000-00007.

Abstract

Recently, the American National Standards Institute (ANSI) has specified (ANSI S3.1-1991) maximum permissible ambient noise levels (MPANLs) allowed in an audiometric test room that would produce negligible masking of pure tones presented down to O-dB HL for two test conditions (ears covered, ears not covered) and three test frequency ranges (125-8000 Hz, 250-8000 Hz, 500-8000 Hz). In the present study, octave band ambient noise sound pressure levels were measured in 136 audiometric test rooms used for clinical audiometry and compared with the ANSI octave band MPANLs for each test condition and frequency range. For the ears covered condition, about 50% of the rooms were in compliance (passed) with the MPANLs for the 125-8000 Hz and 250-8000 Hz range and 82% for the 500-8000 Hz range. For the ears not covered condition, only about 14% of the rooms passed for the 125-8000 Hz and 250-8000 Hz range and 37% passed for the 500-8000 Hz range. The primary cause of the low pass rate for each test condition and frequency range was attributed to low frequency (125-500 Hz) ambient noise presumably from the ventilation system.

摘要

最近,美国国家标准学会(ANSI)已经规定了(ANSI S3.1 - 1991)听力测试室中允许的最大环境噪声水平(MPANLs),在两种测试条件(耳朵覆盖、耳朵未覆盖)和三个测试频率范围(125 - 8000赫兹、250 - 8000赫兹、500 - 8000赫兹)下,该噪声水平对低至0分贝听力级(HL)的纯音产生的掩蔽效应可忽略不计。在本研究中,测量了136个用于临床听力测试的听力测试室中的倍频程带环境噪声声压级,并将其与每种测试条件和频率范围下的ANSI倍频程带MPANLs进行比较。对于耳朵覆盖的条件,约50%的房间在125 - 8000赫兹和250 - 8000赫兹范围内符合(通过)MPANLs,在500 - 8000赫兹范围内为82%。对于耳朵未覆盖的条件,在125 - 8000赫兹和250 - 8000赫兹范围内只有约14%的房间通过,在500 - 8000赫兹范围内为37%。每种测试条件和频率范围通过率低的主要原因被认为是可能来自通风系统的低频(125 - 500赫兹)环境噪声。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验