[Evaluation of reproducibility and precision of ultrasonic measurements of the calcaneus for diagnosing osteopenia].

作者信息

Korczyk P, Hoszowski K, Tałajko A, Lorenc R S

机构信息

II Oddziału Wewnetrznego Szpitala Kolejowego, Warszawie.

出版信息

Pol Tyg Lek. 1993 Nov;48 Suppl 3:28-30.

PMID:8309833
Abstract

The study aimed at evaluating precision of ultrasound measurements of mineral bone density of the calcaneus, and evaluating correlation between the obtained results with those measured with single photon absorptiometry (SPA) and DEXA in different skeletal sites. In 24 healthy individuals and 44 patients with osteoporosis SOS, BUA, calcaneus stiffness index, spine bone mineral density (L2-L4), femoral neck and calcaneal bone mineral densities were measured with DEXA technique whereas BMD in distal radial bone was measured with SPA. Very good precision of ultrasound measurements--better for SOS (CV = 0.12%), worse for BUA (CV = 1.64%) due to repositioning--was achieved. Statistically significant correlation between ultrasound and DEXA measurements of BMD in calcaneus bone (r = 0.90), and significant but worse correlations with BMD measurements for the spine (r = 0.43), femoral bone neck (r = 0.53), and 1/3 radius, (r = 0.51) were achieved. Ultrasound measurements of bone mineral density have low error of repeatability and good accuracy, enabling to discriminate healthy individuals from osteoporosis patients.

摘要

该研究旨在评估超声测量跟骨骨密度的精度,并评估所获结果与不同骨骼部位单光子吸收测定法(SPA)和双能X线吸收法(DEXA)测量结果之间的相关性。对24名健康个体和44名骨质疏松症患者,采用DEXA技术测量跟骨的声速(SOS)、宽带超声衰减(BUA)、跟骨硬度指数、脊柱骨密度(L2-L4)、股骨颈和跟骨骨密度,而采用SPA测量桡骨远端骨密度。超声测量具有非常好的精度——SOS精度更好(变异系数CV = 0.12%),由于重新定位BUA精度稍差(CV = 1.64%)。跟骨骨密度的超声测量与DEXA测量之间存在统计学显著相关性(r = 0.90),与脊柱(r = 0.43)、股骨颈(r = 0.53)和桡骨1/3处(r = 0.51)的骨密度测量也存在显著但稍弱的相关性。骨密度的超声测量具有低重复性误差和良好的准确性,能够区分健康个体和骨质疏松症患者。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验