开发一种用于哮喘诊断的精确便携式记录峰值流量计。

Development of an accurate portable recording peak-flow meter for the diagnosis of asthma.

作者信息

Hitchings D J, Dickinson S A, Miller M R, Fairfax A J

机构信息

School of Engineering, Staffordshire University, Beaconside, Stafford, UK.

出版信息

J Biomed Eng. 1993 May;15(3):188-92. doi: 10.1016/0141-5425(93)90114-e.

Abstract

This article describes the systematic design of an electronic recording peak expiratory flow (PEF) meter to provide accurate data for the diagnosis of occupational asthma. Traditional diagnosis of asthma relies on accurate data of PEF tests performed by the patients in their own homes and places of work. Unfortunately there are high error rates in data produced and recorded by the patient, most of these are transcription errors and some patients falsify their records. The PEF measurement itself is not effort independent, the data produced depending on the way in which the patient performs the test. Patients are taught how to perform the test giving maximal effort to the expiration being measured. If the measurement is performed incorrectly then errors will occur. Accurate data can be produced if an electronically recording PEF instrument is developed, thus freeing the patient from the task of recording the test data. It should also be capable of determining whether the PEF measurement has been correctly performed. A requirement specification for a recording PEF meter was produced. A commercially available electronic PEF meter was modified to provide the functions required for accurate serial recording of the measurements produced by the patients. This is now being used in three hospitals in the West Midlands for investigations into the diagnosis of occupational asthma. In investigating current methods of measuring PEF and other pulmonary quantities a greater understanding was obtained of the limitations of current methods of measurement, and quantities being measured.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

本文介绍了一种电子记录呼气峰值流速(PEF)仪的系统设计,旨在为职业性哮喘的诊断提供准确数据。哮喘的传统诊断依赖于患者在自己家中和工作场所进行的PEF测试的准确数据。不幸的是,患者产生和记录的数据存在很高的错误率,其中大多数是转录错误,有些患者还会伪造记录。PEF测量本身并非不依赖于用力程度,所产生的数据取决于患者进行测试的方式。会教导患者如何进行测试,在测量呼气时尽最大努力。如果测量执行不正确,就会出现误差。如果开发出一种电子记录PEF仪器,就能产生准确的数据,从而使患者无需承担记录测试数据的任务。它还应该能够确定PEF测量是否正确执行。制定了记录PEF仪的需求规格说明书。对一种市售电子PEF仪进行了改装,以提供准确连续记录患者所产生测量值所需的功能。目前西米德兰兹郡的三家医院正在使用该仪器来研究职业性哮喘的诊断。在研究当前测量PEF和其他肺功能指标的方法时,对当前测量方法的局限性以及所测量的指标有了更深入的了解。(摘要截选于250字)

相似文献

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验