Boone J M, Pfeiffer D E, Strauss K J, Rossi R P, Lin P J, Shepard J S, Conway B J
Department of Radiology, University of California Davis, Sacramento 95817.
Med Phys. 1993 May-Jun;20(3):789-94. doi: 10.1118/1.597033.
Fluoroscopic procedures, in general, result in much higher exposures to patients than do most types of radiographic procedures [National Council on Radiation Protection, Report 100, p. 31 (1989)]. In spite of this, fluoroscopic exposure rates can vary widely between systems, and often for no apparent reason. The charge of AAPM Task Group No. 11 was to evaluate fluoroscopic exposure rates at the entrant surface of the x-ray image intensifier, and to disseminate this information so that medical physicists could compare their own exposure rate measurements with typical values. The measurement protocol was defined for various system configurations. Sheets of copper were used to attenuate the x-ray beam, and the input exposure rate at the image intensifier (at the input mode closest to 23-cm diameter) in the absence of a scattering medium was determined. With 2 mm of copper as x-ray beam filtration, the median fluoroscopic exposure rate at the image intensifier was found to be 16.5 nC/kg/s (64.0 microR/s), with an average kV of 77 and mA of 2.0 (n = 62).
一般来说,透视检查给患者带来的辐射剂量比大多数类型的放射检查要高得多[国家辐射防护委员会,第100号报告,第31页(1989年)]。尽管如此,不同系统之间的透视辐射剂量率差异很大,而且往往没有明显原因。美国医学物理学家协会第11特别工作组的任务是评估X射线影像增强器入口表面的透视辐射剂量率,并传播这些信息,以便医学物理学家能够将自己的辐射剂量率测量值与典型值进行比较。针对各种系统配置定义了测量方案。使用铜片来衰减X射线束,并在没有散射介质的情况下确定影像增强器处的输入辐射剂量率(在最接近23厘米直径的输入模式下)。以2毫米铜作为X射线束滤过,发现影像增强器处的透视辐射剂量率中位数为16.5纳库仑/千克/秒(64.0微伦琴/秒),平均千伏数为77,毫安数为2.0(n = 62)。