Graves V B, Rappe A H, Smith T P, Sepetka I, Ahuja A, Strother C M
Department of Radiology, University of Wisconsin Clinical Science Center, Madison 53792-0001.
AJNR Am J Neuroradiol. 1993 Jul-Aug;14(4):804-8.
A new retrieval device for use in small vessels was evaluated in a canine model. Retrieval was attempted on 25 coils, three silicone balloons, and five catheter and guide wire fragments placed in the renal, pulmonary, hepatic, brachiocephalic, subclavian, carotid, and maxillary arteries. This was successful in 91% (30/33). The snare could also be selectively placed in the vertebral and proximal basilar arteries.
一种用于小血管的新型取出装置在犬类模型中进行了评估。对放置在肾动脉、肺动脉、肝动脉、头臂动脉、锁骨下动脉、颈动脉和上颌动脉中的25个线圈、3个硅胶球囊以及5个导管和导丝碎片尝试进行取出操作。成功率为91%(33例中的30例)。圈套器也可选择性地放置在椎动脉和基底动脉近端。