Lang A, Portney V
Allergen Therapeutics, Surgical RD&E, Irvine, California 92718-2020.
J Cataract Refract Surg. 1993 Jul;19(4):505-12. doi: 10.1016/s0886-3350(13)80615-3.
The optical performance of new multifocal intraocular lens designs is frequently assessed using the modulation transfer function (MTF). We discuss the relationship between the MTF and clinical measures of human visual function, such as threshold visual acuity and contrast sensitivity. Using in vitro MTF measurements of a human eye model containing a multifocal or monofocal intraocular lens, we predict relative changes in acuity and contrast sensitivity and outline the techniques using a simple model of human retinal threshold detection. Specific concepts introduced include the visual acuity graph, predicted visual acuity graph, and predicted contrast sensitivity function.
新型多焦点人工晶状体设计的光学性能常通过调制传递函数(MTF)进行评估。我们讨论了MTF与人类视觉功能临床指标之间的关系,如阈值视力和对比敏感度。通过对包含多焦点或单焦点人工晶状体的人眼模型进行体外MTF测量,我们预测了视力和对比敏感度的相对变化,并使用简单的人类视网膜阈值检测模型概述了相关技术。引入的具体概念包括视力图、预测视力图和预测对比敏感度函数。