Bond V, Bassett D R, Howley E T, Lewis J, Walker A J, Swan P D, Tearney R J, Adams R G
Department of Human Performance and Sport Studies, University of Tennessee, Knoxville.
Br J Sports Med. 1993 Jun;27(2):107-9. doi: 10.1136/bjsm.27.2.107.
The application of automated blood pressure measurement during exercise has been limited by inaccuracies introduced by the effects of accompanying motion and noise. We evaluated a newly developed automated blood pressure monitor for measuring exercise blood pressure (Colin STBP-680; Colin, San Antonio, Texas, USA). The STBP-680 uses acoustic transduction with the assistance of the electrocardiogram R-wave to trigger the sampling period for blood pressure measurement. The automated monitor readings were compared with simultaneous technician mercury sphygmomanometric readings in the same arm. Blood pressure was measured in 18 men at rest and during exercise at 40% VO2 peak, (low intensity), 70% VO2 peak (moderate intensity) and VO2 peak (high intensity) on the cycle ergometer. Mean(s.d.) systolic blood pressure difference between the automated monitor and mercury manometer readings at rest and during exercise at low, moderate and high work intensities were 3(0) mmHg, 3(2) mmHg, 1(1) mmHg, and 0(11) mmHg respectively (analysis of variance; P > 0.05). Resting diastolic blood pressure obtained with the STBP-680 was similar to the mercury manometer readings (78(10) versus 81(7) mmHg (P > 0.05). Exercise diastolic pressure at the low level of work intensity was almost identical between the automated monitor and mercury manometer readings (64(8) versus 65(10) mmHg (not significant)). Diastolic blood pressure readings between the STBP-680 and mercury manometer showed a greater difference at the moderate and high workloads (11 mmHg and 9 mmHg, respectively), but this difference was not significant (P > 0.05).(ABSTRACT TRUNCATED AT 250 WORDS)
运动期间自动血压测量的应用受到伴随运动和噪音影响所导致的误差限制。我们评估了一种新开发的用于测量运动血压的自动血压监测仪(科林STBP - 680;科林公司,美国得克萨斯州圣安东尼奥)。STBP - 680利用声学传导并借助心电图R波来触发血压测量的采样周期。将自动监测仪的读数与同一手臂同时由技术人员使用汞柱式血压计测得的读数进行比较。对18名男性在静息状态下以及在功率自行车上进行40%最大摄氧量(低强度)、70%最大摄氧量(中等强度)和最大摄氧量(高强度)运动期间的血压进行了测量。自动监测仪与汞柱式血压计在静息状态以及低、中、高工作强度运动期间测得的平均(标准差)收缩压差值分别为3(0)mmHg、3(2)mmHg、1(1)mmHg和0(1)mmHg(方差分析;P>0.05)。使用STBP - 680测得的静息舒张压与汞柱式血压计读数相似(78(10)对81(7)mmHg(P>0.05))。在低工作强度运动时,自动监测仪与汞柱式血压计的运动舒张压读数几乎相同(64(8)对65(10)mmHg(无显著差异))。在中等和高工作负荷下,STBP - 680与汞柱式血压计的舒张压读数差异更大(分别为11 mmHg和9 mmHg),但该差异不显著(P>0.05)。(摘要截选至250字)