电子根尖定位仪的准确性:对上颌磨牙的临床评估

Accuracy of an electronic apex locator: a clinical evaluation in maxillary molars.

作者信息

Hembrough J H, Weine F S, Pisano J V, Eskoz N

机构信息

Department of Endodontics, Loyola University School of Dentistry, Maywood, IL 60153.

出版信息

J Endod. 1993 May;19(5):242-6. doi: 10.1016/S0099-2399(06)81300-2.

Abstract

Identifying the point of apical termination in endodontic therapy is of indisputable importance. In this study, the accuracy of an electronic apex locator was clinically evaluated in 26 maxillary molar teeth, and comparisons were made to the radiographic and direct anatomical measurements of tooth length. The results indicated that a higher percentage of acceptable measurements were obtained by the radiographic method as compared with the electronic method (88.5% to 73.1%). Additionally, inaccurate electronic estimates were long in every instance. The results of this study do not support the replacement of radiographs with electronic devices in endodontic therapy.

摘要

确定牙髓治疗中根尖终止点具有无可争议的重要性。在本研究中,对26颗上颌磨牙的电子根尖定位仪的准确性进行了临床评估,并与牙齿长度的影像学测量和直接解剖测量进行了比较。结果表明,与电子方法相比,影像学方法获得的可接受测量比例更高(88.5%对73.1%)。此外,在每种情况下,电子估计值不准确的情况都表现为过长。本研究结果不支持在牙髓治疗中用电子设备取代X线片。