Multilayer film analysis: evaluation of ion-selective electrolyte slides.

作者信息

Costello P, Kubasik N P, Brody B B, Sine H E, Bertsch J A, D'Souza J P

出版信息

Clin Chem. 1983 Jan;29(1):129-32.

PMID:6401231
Abstract

We evaluated the Kodak Ektachem multilayer ion-selective electrolyte slides. For various types and concentrations of control material the precision (CV) within- and between-day ranged from 0.5 to 1.3% (1.7-2.1%) for sodium, 1.2 to 2.2% (2.7-2.9%) for potassium, 2.9 to 4.6% (5.9-6.7%) for carbon dioxide, and 0.7 to 1.6% (1.3-1.4%) for chloride. For all these analytes, analytical recovery was about 100%, except in the supra-physiological ranges, for which carbon dioxide recovery was about 110-120%. Either serum or heparin-treated plasma can be used, interchangeably, for analysis; use of serum treated with lithium iodoacetate is unacceptable. Comparisons with results by continuous-flow procedures demonstrated good correlation for sodium, potassium, and chloride; carbon dioxide comparisons indicate an Ektachem calibrator change may be required. Abnormally low protein concentrations or lipemia had no observed effects on results for electrolytes. Abnormally high protein concentrations affect sodium results slightly (approximately 5 mmol/L).

摘要

我们评估了柯达Ektachem多层离子选择性电解质载片。对于各种类型和浓度的对照物质,钠的日内和日间精密度(CV)范围为0.5%至1.3%(1.7% - 2.1%),钾为1.2%至2.2%(2.7% - 2.9%),二氧化碳为2.9%至4.6%(5.9% - 6.7%),氯为0.7%至1.6%(1.3% - 1.4%)。对于所有这些分析物,除了在超生理范围内二氧化碳回收率约为110% - 120%外,分析回收率约为100%。血清或肝素处理的血浆均可用于分析,二者可互换使用;使用碘乙酸锂处理的血清是不可接受的。与连续流动法结果的比较表明,钠、钾和氯具有良好的相关性;二氧化碳比较表明可能需要更换Ektachem校准物。异常低的蛋白质浓度或脂血对电解质结果无观察到的影响。异常高的蛋白质浓度对钠结果有轻微影响(约5 mmol/L)。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验