Oswald M H, Jakob R P, Schneider E, Hoogewoud H M

机构信息

From M.E.M.-Institute for Biomechanics, University of Bern, Switzerland.

出版信息

J Arthroplasty. 1993 Aug;8(4):419-26. doi: 10.1016/s0883-5403(06)80042-2.

Abstract

Alignment of the anatomical and mechanical axes of the lower extremities was analyzed using standardized radiographs in two series of 66 femur and 44 tibia specimens. A new anatomical axis, the distal femoral anatomical axis, was introduced. This axis corresponds to the overall femoral anatomical axis and encloses a valgus angle of 6 degrees with the mechanical axis of the femur. It allows for design of shorter intramedullary guiding instruments that will still be in congruency with true femur geometry. Comparison with computerized calculations of angles on computed tomography scans on a series of 38 bones shows a high precision of measurement on standard radiographs in neutral rotation.

摘要

使用标准化X线片对两个系列的66个股骨和44个胫骨标本的下肢解剖轴和机械轴进行了对齐分析。引入了一种新的解剖轴,即股骨远端解剖轴。该轴与整个股骨解剖轴相对应,与股骨机械轴形成6度的外翻角。它允许设计更短的髓内导向器械,这些器械仍将与真实的股骨几何形状一致。与一系列38块骨骼的计算机断层扫描上的角度计算机计算结果相比,显示在中立旋转时标准X线片上的测量具有高精度。