Abstract

To estimate the stray radiation exposure which will be delivered to the doctor during a coronary angiography we made phantom measurements. The dependence of the stray radiation from the projection angle as well as from the tube tension for fluoroscopy and cinematography was measured in the three space coordinates. Out of these results the exposure per examination can be estimated for each place and hight of the doctor. As an example a determination of exposure under practical condition as well as the means of radiation protection are discussed.

摘要

为了估算在冠状动脉造影过程中传递给医生的杂散辐射剂量,我们进行了体模测量。在三个空间坐标中测量了透视和摄影时杂散辐射与投影角度以及管电压的关系。根据这些结果,可以估算出医生在每个位置和高度时每次检查的辐射剂量。作为示例,讨论了实际条件下辐射剂量的测定以及辐射防护方法。