Characterization of a multi-leaf collimator system.

作者信息

Galvin J M, Smith A R, Lally B

机构信息

New York University Medical Center, Tisch Hospital, Division of Radiation Oncology, NY 10016.

出版信息

Int J Radiat Oncol Biol Phys. 1993 Jan 15;25(2):181-92. doi: 10.1016/0360-3016(93)90339-w.

Abstract

Commissioning measurements for a multi-leaf collimator installed on a dual energy accelerator with 6 and 15 MV photons are described. Detailed dosimetric characterization of the multi-leaf collimator is a requirement for modeling the collimator with treatment planning software. Measurements include a determination of the penumbra width, leaf transmission, between-leaf leakage, and localization of the leaf ends and sides. Standard radiographic film was used for the penumbra measurements, and separate experiments using radiochromic film and thermoluminescent dosimeters were performed to verify that distortions of the dose distribution at an edge due to changing energy sensitivity of silver bromide film are negligible. Films were analyzed with a scanning laser densitometer with a 210 micron spot. Little change in the penumbra edge distribution was noted for different positions of a leaf in the field. Experiments localizing the physical end of the leaves showed less than 1 mm deviation from the 50% decrement line. This small difference is attributed to the shaped end on the leaves. One side of a single leaf corresponded to the 50% decrement line, but the opposite face was aligned with a lower value. This difference is due to the tongue and groove used to decrease between-leaf leakage. For both energies, approximately 2% of photons incident on the multi-leaf collimator are transmitted and an additional 0.5% leakage occurs between the leaves. Alignment of the leaves to form a straight edge results in a penumbra profile which compares favorably with the standard technique of using alloy blocks. When the edge is stepped, the isodose lines follow the leaf pattern and the boundary is poorly defined compared to divergent blocks.

摘要

描述了安装在具有6和15兆伏光子的双能加速器上的多叶准直器的调试测量。多叶准直器的详细剂量学特征是使用治疗计划软件对准直器进行建模的要求。测量包括半值层宽度的测定、叶片透射率、叶片间泄漏以及叶片端部和侧面的定位。使用标准的射线照相胶片进行半值层测量,并进行单独的实验,使用放射变色胶片和热释光剂量计来验证由于溴化银胶片能量灵敏度变化而导致的边缘剂量分布畸变可以忽略不计。用光斑为210微米的扫描激光密度计分析胶片。对于叶片在射野中的不同位置,半值层边缘分布变化不大。确定叶片物理端部的实验表明,与50%剂量递减线的偏差小于1毫米。这种小差异归因于叶片上的成型端部。单个叶片的一侧对应于50%剂量递减线,但相对的面与较低的值对齐。这种差异是由于用于减少叶片间泄漏的榫槽结构。对于两种能量,入射到多叶准直器上的光子中约2%被透射,叶片间额外发生0.5%的泄漏。叶片对齐形成直边时,半值层轮廓与使用合金块的标准技术相比具有优势。当边缘呈阶梯状时,等剂量线遵循叶片模式,与发散块相比,边界定义不清。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验