Improved accuracy of indirect blood pressure measurement in patients with obese arms.

作者信息

Stolt M, Sjönell G, Aström H, Rössner S, Hansson L

机构信息

Department of Development of Primary Health Care, Stockholm, Sweden.

出版信息

Am J Hypertens. 1993 Jan;6(1):66-71. doi: 10.1093/ajh/6.1.66.

Abstract

We studied 20 patients with moderate to severe obesity (body mass index 37 +/- 8 kg/m2) with obese arms (arm circumference 37 +/- 4 cm). Their blood pressure was measured directly in the brachial artery in one arm and simultaneously indirectly in the other arm using either a large standard cuff (rubber bag 12 x 35 cm) or a new cuff (the Tricuff, Pressure Group AB, Stockholm) containing three rubber bags of different sizes, which automatically selected the appropriately sized bag in relation to arm circumference. Both cuffs showed a significant overestimation of the diastolic blood pressure (standard cuff 13 mm Hg, P < .001, Tricuff 6 mm Hg, P < .01). The error of the standard cuff was significantly greater than that of the Tricuff (P < .001). The differences in systolic blood pressure between the intraarterially and the indirectly measured results were small (0.2 to 3 mm Hg) and not statistically significant with either cuff. The new Tricuff offered an advantage as compared to a large standard cuff in terms of a lesser overestimation of the diastolic blood pressure. In practical terms, this should lead to a reduction in the overestimation of diastolic hypertension in obese individuals.

摘要

我们研究了20例中度至重度肥胖患者(体重指数37±8kg/m²),其手臂肥胖(臂围37±4cm)。在其中一只手臂的肱动脉直接测量血压,同时在另一只手臂使用大型标准袖带(橡胶袋12×35cm)或一种新袖带(Tricuff,斯德哥尔摩压力集团AB公司)间接测量血压,新袖带包含三个不同尺寸的橡胶袋,可根据臂围自动选择尺寸合适的袋子。两种袖带均显示出对舒张压的显著高估(标准袖带高估13mmHg,P<.001;Tricuff高估6mmHg,P<.01)。标准袖带的误差显著大于Tricuff袖带(P<.001)。动脉内测量与间接测量结果之间的收缩压差异较小(0.2至3mmHg),使用两种袖带时均无统计学意义。与大型标准袖带相比,新型Tricuff袖带在舒张压高估方面具有优势。实际上,这将减少肥胖个体中舒张期高血压的高估情况。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验