Draeger J, Wiezorrek R, Hock B, Gerhard E, Klemm M
Universitäts-Augenklinik Hamburg.
Ophthalmologe. 1993 Feb;90(1):51-3.
The microprocessor-controlled applanation tonometer was calibrated in force measurement and area detection by using weights and silicon calibration pistons. This makes it possible to compare the technical precision of the instrument under standardized conditions and the results of previous clinical examinations. With the instrument in a vertical position, the force measurements meets the official calibration requirement. Force measurements in a horizontal position will follow in order to demonstrate that exact measurements can be taken with the instrument irrespective of position. A suitable method for calibrating the optical sensor by using the silicone pistons for area detection is shown. This method is to be extended by using calibration pistons of different diameters, which are within the permitted calibration tolerances.
通过使用砝码和硅校准活塞,对微处理器控制的压平眼压计进行了力测量和面积检测校准。这使得在标准化条件下比较该仪器的技术精度和先前临床检查结果成为可能。仪器处于垂直位置时,力测量符合官方校准要求。接下来将进行水平位置的力测量,以证明无论仪器处于何种位置都能进行精确测量。展示了一种使用硅活塞进行面积检测来校准光学传感器的合适方法。该方法将通过使用不同直径且在允许校准公差范围内的校准活塞来扩展。