McManus C D, Neubert K D, Cramer E
Physikalisch-Technische Bundesanstalt, Institut Berlin, Germany.
Comput Biomed Res. 1993 Feb;26(1):48-67. doi: 10.1006/cbmr.1993.1003.
Diverse digital methods have been advanced previously to remove power line (AC) interference in the ECG. Representative notch filters, adaptive filters and a globally derived filter are surveyed in this study; their performances are compared on artificial signals as well as actual ECGs. The ECGs, recorded at four European medical centers, are from the Common Standards in Electrocardiography (CSE) ECG tape library. AC interference in these ECGs is shown to exhibit two qualities especially relevant to filter design: considerable deviations from a nominal 50 Hz frequency and substantial noise at higher harmonics. Some criteria and useful quantitative measures are suggested to evaluate AC digital filters.
此前已经提出了多种数字方法来消除心电图(ECG)中的电源线(AC)干扰。本研究对代表性的陷波滤波器、自适应滤波器和全局推导滤波器进行了调查;在人工信号以及实际心电图上对它们的性能进行了比较。在四个欧洲医疗中心记录的心电图来自心电图通用标准(CSE)心电图磁带库。这些心电图中的交流干扰表现出与滤波器设计特别相关的两个特性:与标称50Hz频率有相当大的偏差以及高次谐波处有大量噪声。提出了一些标准和有用的定量措施来评估交流数字滤波器。