Belin M W
Albany Medical College Department of Ophthalmology, New York 12203.
J Cataract Refract Surg. 1993;19 Suppl:188-92.
The PAR Corneal Topography System (CTS) is a computer-driven corneal imaging system that uses close-range raster photogrammetry to measure and produce a topographic map of the corneal surface. The CTS determines distortion in a projected two-dimensional grid. Unlike Placido-disc-based videokeratoscopes, the PAR CTS produces a true topographic map (elevation map) and requires neither a smooth reflective surface nor precise spatial alignment for accurate imaging. Because the system uses two noncoaxial optical paths, it can be integrated into other optical devices. A modified CTS was integrated into an experimental erbium: YAG photoablative laser. The CTS successfully imaged corneas before, after, and during laser photoablation. Its ability to image nonreflective surfaces and to be integrated into other optical systems may make it suitable for intraoperative refractive monitoring.
PAR角膜地形图系统(CTS)是一种计算机驱动的角膜成像系统,它使用近程光栅摄影测量法来测量并生成角膜表面的地形图。CTS可确定投射二维网格中的畸变。与基于普拉西多盘的视频角膜镜不同,PAR CTS可生成真正的地形图(高度图),并且精确成像既不需要光滑的反射表面,也不需要精确的空间对准。由于该系统使用两条非同轴光路,因此可以集成到其他光学设备中。一种改良的CTS被集成到一台实验性铒:钇铝石榴石光烧蚀激光器中。CTS成功地对激光光烧蚀前、后及过程中的角膜进行了成像。其对非反射表面成像以及集成到其他光学系统中的能力可能使其适用于术中屈光监测。