Wülker N, Siebert W E
Orthopädische Klinik, Medizinischen Hochschule Hannover.
Aktuelle Radiol. 1993 Mar;3(2):100-4.
22 patients suffering from different orthopaedic disease patterns were x-rayed with direct x-ray magnification with focus diameters between 0.012 and 0.100 mm. This yielded improved resolution of detail structures, especially in tumours and aseptic bone necroses. Due to the relatively low efficiency of the x-ray tube (conditioned by the physics of the x-ray equipment), however, exposure times of several seconds were required, and the voltage had to be higher than in conventional x-ray procedure. This resulted in lack of sharpness due to movement and to a loss in contrast. Hence, the clinical use of focus diameters below 0.1 mm does not seem to be meaningful.
对22名患有不同骨科疾病类型的患者进行了直接X射线放大摄影,焦点直径在0.012至0.100毫米之间。这提高了细节结构的分辨率,尤其是在肿瘤和无菌性骨坏死方面。然而,由于X射线管的效率相对较低(受X射线设备物理特性限制),需要数秒的曝光时间,并且电压必须高于传统X射线检查程序。这导致了因移动而产生的图像模糊以及对比度的损失。因此,临床使用焦点直径低于0.1毫米似乎并无意义。