[Inhaled nitric monoxide. Application and continuous measurement of concentration].

作者信息

Benzing A, Beyer U, Kiefer P, Geiger K

机构信息

Anaesthesiologische Universitätsklinik Freiburg.

出版信息

Anaesthesist. 1993 Mar;42(3):175-8.

PMID:8480905
Abstract

Inhaled nitric oxide (NO) is a selective pulmonary vasodilator that may be useful in the treatment of patients with severe pulmonary hypertension. We describe a delivery system of inhaled NO that allows safe application and continuous measurement of the inspired NO concentration during mechanical ventilation. From a gas cylinder containing NO in N2 (600 ppm NO), an adjustable amount of gas is introduced into the inspiratory side of the tubing system via a pressure reduction valve, a magnetic valve, and a special injection nozzle. The NO concentration is diluted to the desired value by the tidal volume. The magnetic valve is connected to the ventilator and opens at the beginning of each inspiration and closes after a predetermined time. The gas volume is proportional to the pressure at the magnetic valve and the opening time. To monitor the inspiratory NO concentration, a specimen of gas is taken from an angle-connector and passed over an electrochemical sensor. The second nozzle of the sensor is connected to a water seal, which is adjusted to the positive end-expiratory pressure level of the ventilator to insure that the gas flow over the sensor is limited to inspiration.

摘要

吸入一氧化氮(NO)是一种选择性肺血管扩张剂,可能对治疗重度肺动脉高压患者有用。我们描述了一种吸入NO输送系统,该系统可在机械通气期间安全应用并持续测量吸入的NO浓度。从装有N₂中含NO(600 ppm NO)的气瓶中,通过减压阀、电磁阀和特殊喷嘴将可调量的气体引入管路系统的吸气侧。NO浓度通过潮气量稀释至所需值。电磁阀连接到呼吸机,在每次吸气开始时打开,并在预定时间后关闭。气体体积与电磁阀处的压力和打开时间成正比。为监测吸入的NO浓度,从角形连接器采集气体样本并通过电化学传感器。传感器的第二个喷嘴连接到水封,水封调节到呼吸机的呼气末正压水平,以确保传感器上的气流仅限于吸气过程。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验