[机械通气期间一种新型一氧化氮输送系统的评估]

[Evaluation of a new nitric oxide delivery system during mechanical ventilation].

作者信息

Noguchi T, Miyakawa H, Mori M, Kitano T, Iwasaka H, Oda S, Taniguchi K, Honda N

机构信息

Intensive Care Unit, Oita Medical University Hospital.

出版信息

Masui. 1994 Jul;43(7):1083-6.

PMID:7933483
Abstract

A new nitric oxide delivery and continuous monitoring system is described. During mechanical ventilation, this new system connected with Siemens Servo 900C ventilator was shown to be able to provide a constant inspired NO concentration (10-100 ppm) using chemiluminescence technique for NO analysis. Gas was analysed at the mixing chamber in front of the ventilator inlet and inspiratory tube connected with the soda-lime carbon-dioxide absorber. Both NO concentrations showed a good correlation (r = 0.99). The actual NO concentration from the NO supply cylinder was 1154 ppm and NO2 concentration was 14 ppm. In mongrel dogs, after 20 minutes of NO inhalation (10-100 ppm), the blood methemoglobin level reached a peak value of 2.2% starting from the pre-inhalation level of 0%. To optimize the safety of the clinical application of NO, its concentration should be measured continuously with chemiluminescence technique.

摘要

描述了一种新的一氧化氮输送和连续监测系统。在机械通气期间,该新系统与西门子Servo 900C呼吸机相连,结果表明,使用化学发光技术进行一氧化氮分析时,它能够提供恒定的吸入一氧化氮浓度(10 - 100 ppm)。在呼吸机入口前的混合室以及与碱石灰二氧化碳吸收器相连的吸气管处对气体进行分析。两种一氧化氮浓度显示出良好的相关性(r = 0.99)。来自一氧化氮供应气瓶的实际一氧化氮浓度为1154 ppm,二氧化氮浓度为14 ppm。在杂种犬中,吸入一氧化氮(10 - 100 ppm)20分钟后,血液高铁血红蛋白水平从吸入前的0%达到峰值2.2%。为优化一氧化氮临床应用的安全性,应使用化学发光技术连续测量其浓度。

相似文献

5
A safe clinical system for nitric oxide inhalation therapy for pediatric patients.一种用于儿科患者一氧化氮吸入治疗的安全临床系统。
Pediatr Pulmonol. 1996 Sep;22(3):174-81. doi: 10.1002/(SICI)1099-0496(199609)22:3<174::AID-PPUL6>3.0.CO;2-P.
6
Variation of nitric oxide concentration during inspiration.吸气过程中一氧化氮浓度的变化。
Crit Care Med. 1997 Feb;25(2):365-71. doi: 10.1097/00003246-199702000-00028.
8
[In vitro study of the formation of NO2 in inhalation of nitrogen monoxide].[一氧化氮吸入过程中二氧化氮形成的体外研究]
Anasthesiol Intensivmed Notfallmed Schmerzther. 1994 May;29(3):157-62. doi: 10.1055/s-2007-996706.
10
Evaluation of a new system for ventilatory administration of nitric oxide.一种新型一氧化氮通气给药系统的评估
Acta Anaesthesiol Scand. 1993 Oct;37(7):687-91. doi: 10.1111/j.1399-6576.1993.tb03790.x.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验