计算机X线摄影中的散射。

Scatter in computed radiography.

作者信息

Tucker D M, Souto M, Barnes G T

机构信息

Department of Radiology, University of Alabama Hospitals and Clinics, Birmingham 35233.

出版信息

Radiology. 1993 Jul;188(1):271-4. doi: 10.1148/radiology.188.1.8511311.

Abstract

The effects of scatter radiation on images obtained with storage phosphor computed radiography (CR) were evaluated. The scatter degradation factor was extended to account separately for forward scatter and backscatter. Values for the ratios of forward scatter to primary radiation (sf/p) and backscatter to primary radiation (sb/p) were measured for various x-ray tube potentials, phantom thicknesses, and field sizes for CR imaging plates (IPs). Similar measurements were made for a common screen-film system. The sb/p values for the IPs were 0.05-3.80. Standard CR cassettes do not have lead foil behind the IP to control backscatter; the results of this study indicate that such control is needed. The sf/p values were 40%-70% higher in IPs (50%-120% higher when backscatter was included) than those in the conventional screen-film system. This difference is attributed to the differences in the K-absorption edges of barium and gadolinium and the fact that scatter has lower effective energy than the primary beam as they exit the patient.

摘要

评估了散射辐射对使用存储磷光体计算机放射摄影(CR)获得的图像的影响。散射降解因子被扩展,以便分别考虑前向散射和反向散射。针对CR成像板(IP)的各种X射线管电位、体模厚度和野大小,测量了前向散射与一次辐射的比值(sf/p)以及反向散射与一次辐射的比值(sb/p)。对一种常见的屏-片系统进行了类似测量。IP的sb/p值为0.05 - 3.80。标准CR暗盒在IP后面没有铅箔来控制反向散射;本研究结果表明需要这种控制。IP的sf/p值比传统屏-片系统高40% - 70%(当包括反向散射时高50% - 120%)。这种差异归因于钡和钆的K吸收边的差异以及散射在离开患者时具有比一次射线更低的有效能量这一事实。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验