Suppr超能文献

流量触发式婴儿呼吸机的反应。

The response of flow-triggered infant ventilators.

作者信息

Nishimura M, Hess D, Kacmarek R M

机构信息

Respiratory Care Department Laboratory, Massachusetts General Hospital, Boston 02114, USA.

出版信息

Am J Respir Crit Care Med. 1995 Dec;152(6 Pt 1):1901-9. doi: 10.1164/ajrccm.152.6.8520753.

Abstract

Patient-triggered ventilation (PTV) has not been feasible for infants because of large trigger pressures and long delay times with pressure-triggered systems. Recently, four infant ventilators with flow triggering have become available. We questioned if delay times, trigger pressures, and trigger work with these ventilators would be acceptable for PTV in infants. All ventilators were attached via 3-, 4-, and 5-mm endotracheal tubes to a spontaneously breathing infant lung model. The lung simulator was set at an inspiratory time of 0.65 s, tidal volume of 15, 30, and 45 ml, and 0 and 5 cm H2O positive end-expiratory pressure (PEEP). Delay time, trigger pressure, and trigger work were determined from pressure measured at the proximal airway, trachea, and alveolus. There were significant differences between the endotracheal tube sizes, sites of measurement, ventilatory demand and ventilator brand at each PEEP level for delay time, trigger pressure, and trigger work (p < 0.001). Delay time was greatest with the 3-mm endotracheal tube at high ventilatory drive (maximum 138.2 +/- 2.1 ms). Both trigger pressure (minimum 0.23 +/- 0.02 cm H2O) and trigger work (minimum 0.05 +/- 0.01 g.ml) increased with decreasing endotracheal tube size, increasing ventilatory demand, use of PEEP, and site of measurement: alveolus > trachea > airway (maximum: trigger pressure 5.04 +/- 0.02 cm H2O; trigger work 114.48 +/- 0.88 g.ml). PTV may not be appropriate under conditions of increased ventilatory drive and small endotracheal tube size in infants.

摘要

由于压力触发系统的触发压力大且延迟时间长,患者触发通气(PTV)对婴儿并不可行。最近,出现了四款具有流量触发功能的婴儿呼吸机。我们质疑这些呼吸机的延迟时间、触发压力和触发功是否适用于婴儿的PTV。所有呼吸机均通过3毫米、4毫米和5毫米的气管内导管连接到一个自主呼吸的婴儿肺模型上。肺模拟器设置的吸气时间为0.65秒,潮气量为15毫升、30毫升和45毫升,呼气末正压(PEEP)为0厘米水柱和5厘米水柱。延迟时间、触发压力和触发功由在近端气道、气管和肺泡处测得的压力确定。在每个PEEP水平下,气管内导管尺寸、测量部位、通气需求和呼吸机品牌之间的延迟时间、触发压力和触发功存在显著差异(p < 0.001)。在高通气驱动时,3毫米气管内导管的延迟时间最长(最大138.2 +/- 2.1毫秒)。触发压力(最小0.23 +/- 0.02厘米水柱)和触发功(最小0.05 +/- 0.01克·毫升)均随着气管内导管尺寸减小、通气需求增加、PEEP的使用以及测量部位的变化而增加:肺泡>气管>气道(最大值:触发压力5.04 +/- 0.02厘米水柱;触发功114.48 +/- 0.88克·毫升)。在婴儿通气驱动增加和气管内导管尺寸较小的情况下,PTV可能并不适用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验