Gains M J, Grodecki K M, Jacobs R M, Dyson D, Foster R A
Department of Pathology, Ontario Veterinary College, University of Guelph.
Can J Vet Res. 1995 Jul;59(3):238-40.
The precision and accuracy of an indirect oscillometric blood pressure measurement technique (Dinamap 8100) was assessed in 11 anesthetized Beagle dogs weighing 8 to 11.5 kg. Direct blood pressure measurements were made by catheterization of the lingual artery, and simultaneous indirect measurements were determined by placing a cuff over the median artery (midradial area). Blood pressure measurements at 2 different planes of anesthesia (light and deep) were recorded in triplicate. At a light plane of anesthesia, the Dinamap 8100 underestimated diastolic and mean arterial pressure, and at a deep anesthetic plane overestimated systolic pressure. The indirect technique had good repeatability of systolic pressures. Regression analysis for the 2 techniques showed excellent correlation (r = 0.93). The results indicate that the indirect oscillometric blood pressure measurement technique provides a good estimate of systolic, diastolic, and mean arterial pressure in dogs weighing 8-11.5 kg.