Henderson J, Willson K, Jago J R, Whittingham T A
Regional Medical Physics Department, Newcastle General Hospital, Newcastle Upon Tyne, UK.
Ultrasound Med Biol. 1995;21(5):699-705. doi: 10.1016/0301-5629(94)00158-a.
Surveys published up to 1991 have highlighted a steady increase in the acoustic outputs from diagnostic ultrasound equipment. Since 1991 we have made measurements of the maximum peak negative pressure (p-) and spatial peak temporal average intensity (ISPTA) produced by 223 probes from 82 scanning systems in current clinical use in the Northern Region in the UK. Measurements have also been made of the maximum total acoustic power generated by 45 probes from 17 scanners. The results from these measurements are presented in this article and compared to the results of a similar survey of equipment from both the Northern and Wessex Regions in the UK and published in 1991. The comparison shows that measured ISPTA values have increased approximately sixfold in B mode and approximately threefold in colour Doppler mode. Also, measured total acoustic power values have doubled in pulsed Doppler mode. The present survey also draws attention to some particularly high ISPTA values obtained from a number of probes and scanning systems. This survey has shown that measurements of acoustic outputs from diagnostic ultrasound scanners in current clinical use are substantially higher than reported in earlier surveys and, for certain scanners, the acoustic outputs from scanned beam modes of operation can reach levels hitherto only found in pulsed Doppler mode.
截至1991年发表的调查强调了诊断超声设备的声输出量稳步增加。自1991年以来,我们对英国北部地区目前临床使用的82个扫描系统中的223个探头所产生的最大峰值负压(p-)和空间峰值时间平均强度(ISPTA)进行了测量。我们还对17台扫描仪中的45个探头所产生的最大总声功率进行了测量。本文展示了这些测量结果,并与1991年发表的对英国北部和韦塞克斯地区设备的类似调查结果进行了比较。比较结果显示,在B模式下,测量得到的ISPTA值增加了约六倍,在彩色多普勒模式下增加了约三倍。此外,在脉冲多普勒模式下,测量得到的总声功率值增加了一倍。本次调查还提请注意从一些探头和扫描系统获得的一些特别高的ISPTA值。这项调查表明,目前临床使用的诊断超声扫描仪的声输出量测量值大大高于早期调查中报告的值,而且对于某些扫描仪来说,扫描波束操作模式下的声输出量可以达到迄今为止仅在脉冲多普勒模式下才有的水平。