Oriuchi N, Tokunaga M, Suzuki H, Inoue T, Ohnishi Y, Inoue T, Endo K
Department of Nuclear Medicine, Gunma University, School of Medicine.
Nihon Igaku Hoshasen Gakkai Zasshi. 1995 Sep;55(11):757-62.
The aim of this study was to establish a simple and accurate procedure to calculate the unilateral effective renal plasma flow (ERPF) using 99mTc-MAG3 gamma-camera renographyl. Eleven patients with urological disorders were studied with 99mTc-MAG3 to calculate the renal uptake ratio (RUR), which was defined as the ratio of absorption-corrected renal counts within 1-2 min after intravenous injection to injection radioactivity measured with a gamma-camera. We assumed that 99mTc-MAG3 was distributed in the circulation, moved from the circulation to the kidneys, and was excreted solely from the kidneys. We thus adopted an one-compartment model to calculate 99mTc-MAG3 clearance (CLMAG) using RUR. Sequential p-aminohippurate (PAH) clearance study was performed as a gold standard in all the patients to compare CLMAG and PAH clearance. Obtained CLMAG correlated well with PAH clearance, and this one-compartment model seemed appropriate for describing CLMAG. Unilateral ERPF was accurately measured within two minutes after the intravenous injection of 99mTc-MAG3 in conjunction with the renal imaging study. This procedure is simple and reliable, and requires no blood or urine sampling.
本研究的目的是建立一种简单准确的方法,利用99mTc-MAG3γ相机肾图来计算单侧有效肾血浆流量(ERPF)。对11例泌尿系统疾病患者进行了99mTc-MAG3检查,以计算肾摄取率(RUR),其定义为静脉注射后1-2分钟内经吸收校正的肾脏计数与用γ相机测量的注射放射性之比。我们假设99mTc-MAG3分布于循环中,从循环进入肾脏,并仅从肾脏排泄。因此,我们采用单室模型,利用RUR计算99mTc-MAG3清除率(CLMAG)。对所有患者进行连续对氨基马尿酸(PAH)清除率研究作为金标准,以比较CLMAG和PAH清除率。所得CLMAG与PAH清除率相关性良好,该单室模型似乎适合描述CLMAG。在静脉注射99mTc-MAG3后两分钟内,结合肾脏成像研究可准确测量单侧ERPF。该方法简单可靠,无需采集血液或尿液样本。