Soya K, Obata H, Amano S, Miyata K, Tsuru T
Department of Ophthalmology, Kanto Teishin Hospital, Japan.
Nippon Ganka Gakkai Zasshi. 1995 Nov;99(11):1242-7.
Accurate calibration of ablation rate by excimer laser prerequisite for precise photorefractive keratectomy (PRK). When a polymethylmethacrylate (PMMA) plate is ablated by excimer laser, debris is generated, which may affect homogeneity of the laser beam and energy density, and change the ablation rate. In order to clarify the effects of the debris on the ablation rate, we studied the changes in the ablation rate when the debris was removed by blowing or aspirating over the ablation area during PRK.PMMA plates were ablated using a scanning excimer laser system (EC-5000, NIDEK, Japan) in PRK mode under the following conditions: (1) with air blowing over the ablation area, (2) with aspiration of the debris, and (3) without treatment. The ablation rates were determined by measuring the refractive power of PMMA plates with a lens meter. The ablated surface was observed by scanning electron microscope (SEM). The ablation rate with blowing was the highest among the three conditions, that with aspiration was the second, and that without treatment was the lowest. The ablation rates with blowing showed no significant change when the ablation rates were changed. However, the ablation rates with aspiration or without treatment decreased as the pulse rate increased. The surface ablated during blowing was the smoothest in SEM photographs. We concluded that calibration of the ablation rate using PMMA plates must be done with appropriate air blowing.
准分子激光精确校准消融速率是准分子激光原位角膜磨镶术(PRK)的前提条件。当用准分子激光消融聚甲基丙烯酸甲酯(PMMA)板时,会产生碎片,这可能会影响激光束的均匀性和能量密度,并改变消融速率。为了阐明碎片对消融速率的影响,我们研究了在PRK过程中通过在消融区域吹气或抽吸去除碎片时消融速率的变化。在以下条件下,使用扫描准分子激光系统(EC - 5000,日本尼德克公司)以PRK模式消融PMMA板:(1)在消融区域吹气,(2)抽吸碎片,(3)不进行处理。通过用透镜计测量PMMA板的屈光力来确定消融速率。用扫描电子显微镜(SEM)观察消融表面。在三种条件下,吹气时的消融速率最高,抽吸时次之,不进行处理时最低。当改变消融速率时,吹气时的消融速率没有显著变化。然而,抽吸或不进行处理时的消融速率随着脉冲率的增加而降低。在SEM照片中,吹气时消融的表面最光滑。我们得出结论,使用PMMA板校准消融速率必须在适当吹气的情况下进行。