Sehgal C M, Arger P H, Pugh C R
Hospital of the University of Pennsylvania, Philadelphia, USA.
J Ultrasound Med. 1995 Oct;14(10):735-40. doi: 10.7863/jum.1995.14.10.735.
The aim of this study was to evaluate the influence of postprocessing curves on the visual and quantitative enhancement of image brightness caused by the injection of a contrast agent. A fixed dose (0.45 ml/kg) of contrast agent EchoGen was injected into three dogs, and images of the kidney were investigated using one linear (A) and three nonlinear (B, C, D) postprocessing curves. Curve D gave the best gray scale images and was generally used for diagnostic imaging. Images were analyzed qualitatively in three dogs and quantitatively in one dog for brightness change caused by contrast agent injection. Visually the changes were most identifiable with curve C and least by curve D. Video intensity time curves reflected the qualitative changes. Peak contrast and area under the video density curve changed in the order curve D < curve B approximately curve A < curve C. Choice of the postprocessing curve can have a marked influence on the enhancement of image brightness by a contrast agent. Postprocessing curves optimized to display a wide range of echo levels may not always be an ideal choice for contrast sonography.
本研究的目的是评估后处理曲线对注射造影剂后图像亮度的视觉和定量增强的影响。将固定剂量(0.45 ml/kg)的造影剂EchoGen注入三只犬体内,并使用一条线性(A)和三条非线性(B、C、D)后处理曲线对肾脏图像进行研究。曲线D产生的灰度图像最佳,通常用于诊断成像。对三只犬的图像进行定性分析,对一只犬的图像进行定量分析,以研究注射造影剂引起的亮度变化。视觉上,曲线C引起的变化最明显,曲线D引起的变化最不明显。视频强度时间曲线反映了定性变化。峰值对比度和视频密度曲线下的面积变化顺序为曲线D < 曲线B ≈ 曲线A < 曲线C。后处理曲线的选择对造影剂增强图像亮度有显著影响。为显示广泛回声水平而优化的后处理曲线可能并不总是超声造影的理想选择。