Suppr超能文献

[市售肝素化注射器在血气和电解质测量中的问题]

[Problems of commercially available heparinized syringes in measurements of blood gases and electrolytes].

作者信息

Kinoshita H, Fukuda S, Ataka T, Watanabe I, Shimoji K

机构信息

Department of Anesthesiology, Niigata University School of Medicine.

出版信息

Masui. 1995 Nov;44(11):1563-7.

PMID:8544299
Abstract

We compared the measured values of blood gases and electrolytes by commercially available heparinized syringes (PZ-II, Quick-lite and QS90) with normal and decreased calcium concentrations. With normal calcium concentration, the measured value of [Ca2+] in PZ-II was significantly less than that in Quick-lite and QS90, seemingly due to contained heparin. With hypocalcemia, the measured values of [Ca2+] in Quick-lite and QS90 were significantly greater than those in non-heparinized syringe. The pH and PO2 decreased, and PCO2 increased as time passed. The [Ca2+] was suggested to be more influenced by heparin than by acid-base balance.

摘要

我们比较了市售肝素化注射器(PZ-II、Quick-lite和QS90)在正常钙浓度和降低钙浓度情况下的血气和电解质测量值。在正常钙浓度下,PZ-II中[Ca2+]的测量值显著低于Quick-lite和QS90中的测量值,这似乎是由于其中所含的肝素。在低钙血症情况下,Quick-lite和QS90中[Ca2+]的测量值显著高于非肝素化注射器中的测量值。随着时间的推移,pH值和PO2降低,PCO2升高。[Ca2+]似乎受肝素的影响比受酸碱平衡的影响更大。