Vibration analysis of vessel wall motion with intra vena caval balloon pumping.

作者信息

Shafi A S, Lund L W, Walters F R, Sawzik P J, Reeder G D, Borovetz H S, Hattler B G

机构信息

Department of Surgery, University of Pittsburgh Medical Center, PA 15213-2582, USA.

出版信息

ASAIO J. 1994 Jul-Sep;40(3):M740-2. doi: 10.1097/00002480-199407000-00097.

Abstract

The intravenous membrane oxygenator (IMO) incorporates a centrally positioned balloon surrounded by hollow microporous fibers. Previous studies using this configuration have demonstrated that rhythmic pulsation of the balloon enhances gas exchange, presumably by three dimensional convective mixing. This study sought to characterize vessel wall vibrations imparted by intra vena caval balloon pumping. An in vitro flow loop incorporating a current IMO prototype was used for these measurements. The IMO prototype was inserted in a modeled vena cava on which ultrasonic dimension transducers were mounted on the outer surface. The flow loop was operated at physiologic flow rates. The balloon was activated, and dynamic vessel diameter measurements were recorded as the pumping frequency was varied from 40 to 120 beats per minute (bpm). A Fast Fourier Transform algorithm generated a frequency spectrum at each bpm and for two different balloon configurations; a single balloon versus a tripartite arrangement, the authors' results demonstrate that the mean amplitude of vena caval oscillations varied with bpm, and that this variation followed the trends in oxygen transfer rates. This suggests that the motion of the vessel wall may contribute to convective mixing of blood. In addition, this work demonstrated significant differences in the frequency spectra associated with our two balloon configurations.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验