Acute in vivo studies of the Pittsburgh intravenous membrane oxygenator.

作者信息

Macha M, Federspiel W J, Lund L W, Sawzik P J, Litwak P, Walters F R, Reeder G D, Borovetz H S, Hattler B G

机构信息

Department of Surgery, University of Pittsburgh, Pennsylvania 15219, USA.

出版信息

ASAIO J. 1996 Sep-Oct;42(5):M609-15. doi: 10.1097/00002480-199609000-00060.

Abstract

The efficacy of an innovative intravenous membrane oxygenator (IMO) was tested acutely (6-8 hrs) in seven calves. The IMO prototypes consisted of a central polyurethane balloon within a bundle of hollow fibers with a membrane surface area of 0.14 m2. The IMO devices were inserted through the external jugular vein into the inferior vena cava of anesthetized calves (68.9 +/- 2.3 kg). Rhythmic balloon pulsation (60-120 bpm) was controlled with an intra-aortic balloon pump console. Oxygen sweep gas was delivered through the device at 3.0 L/min. Gas concentrations were monitored continuously by mass spectroscopy. The principal results were as follows: 1) oxygen and carbon dioxide exchange ranged from 125 to 150 ml/min/m2 and 150 to 200 ml/min/m2, respectively; 2) there was at least a 30-50% augmentation of gas exchange with balloon pulsation; 3) maximum exchange occurred with 60-90 bpm balloon pulsations; and 4) hemodynamic parameters remained unchanged. There were no device related complications, and the feasibility of insertion of the device by a cervical cut-down was established. These acute in vivo experiments show that the Pittsburgh IMO device can exchange oxygen and carbon dioxide gases in vivo at levels consistent with this current prototype design, and that intravenous balloon pulsation significantly enhances gas exchange without causing any end-organ damage.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验