Binns S H, Sisson D D, Buoscio D A, Schaeffer D J
Department of Veterinary Clinical Medicine, University of Illinois, Urbana 61801, USA.
J Vet Intern Med. 1995 Nov-Dec;9(6):405-14. doi: 10.1111/j.1939-1676.1995.tb03301.x.
Blood pressure (BP) measurements obtained using 3 indirect BP measuring instruments, a Doppler ultrasonic flowmeter, an oscillometric device, and a photoplethysmograph, were compared with direct arterial pressure measurements in 11 anesthetized cats. The standard deviation of the differences (SDD) between direct and indirect pressure measurements were not significantly different from each other (P < .01), and ranged from 10.82 to 24.32 mm Hg. The high SDD values indicate that indirect BP estimates obtained with all these devices must be interpreted cautiously in individual patients. The mean errors (calculated as the sum of the differences between direct and indirect pressure measurements divided by the number of observations) of the 3 indirect devices were significantly different for systolic (SAP), diastolic (DAP), and mean (MAP) arterial pressures (P < .05). The Doppler and photoplethysmographic devices had the highest overall accuracy, as indicated by mean error values of less than 10 mm Hg. Correlation coefficients varied from .88 to .96 for the Doppler flowmeter, and from .85 to .94 for the photoplethysmograph; for both devices, the regression line slopes were close to unity. The Doppler flowmeter detected a pulse under all experimental conditions. The photoplethysmograph was also efficient in obtaining BP measurements, obtaining over 90% of SAP, DAP, and MAP readings attempted. The oscillometric device was the least accurate, with mean error values varying from 10 to 22 mm Hg. Correlation coefficients were high (.90 to .94) for this device, but the slopes of the regression lines were 0.7 to 0.8, indicating a trend for increased error at higher BP. The oscillometric device tended to underestimate BP by increasing amounts as the BP increased. The oscillometric device was the least efficient device for obtaining BP measurements (P < .01).
在11只麻醉猫身上,将使用3种间接血压测量仪器(多普勒超声流量计、示波装置和光电容积描记仪)测得的血压与直接动脉压测量结果进行了比较。直接和间接血压测量之间差异的标准差(SDD)彼此之间无显著差异(P <.01),范围为10.82至24.32毫米汞柱。较高的SDD值表明,在个体患者中,使用所有这些设备获得的间接血压估计值必须谨慎解读。这3种间接设备在收缩压(SAP)、舒张压(DAP)和平均动脉压(MAP)方面的平均误差(计算方法为直接和间接压力测量之间的差异总和除以观察次数)存在显著差异(P <.05)。多普勒和光电容积描记设备总体准确性最高,平均误差值小于10毫米汞柱。多普勒流量计的相关系数在0.88至0.96之间,光电容积描记仪的相关系数在0.85至0.94之间;对于这两种设备,回归线斜率均接近1。多普勒流量计在所有实验条件下均能检测到脉搏。光电容积描记仪在获取血压测量值方面也很有效,尝试获取的SAP、DAP和MAP读数中超过90%成功获取。示波装置准确性最低,平均误差值在10至22毫米汞柱之间。该装置的相关系数较高(0.90至0.94),但回归线斜率为0.7至0.8,表明在较高血压时误差有增加趋势。示波装置往往随着血压升高而低估血压的幅度越来越大。示波装置是获取血压测量值效率最低的设备(P <.01)。