Suppr超能文献

半搏动式旋转血泵在体外循环中的实验应用。

Experimental use of a semipulsatile rotary blood pump for cardiopulmonary bypass.

作者信息

Mesana T, Morita S, Trinkl J, Demunck J L, Gauthier T, Aucomte F, Havlik P, Montiès J R

机构信息

Laboratory for Surgical Research, University of Aix-Marseilles II, School of Medicine, France.

出版信息

Artif Organs. 1995 Jul;19(7):734-8. doi: 10.1111/j.1525-1594.1995.tb02414.x.

Abstract

We tested our valveless pulsatile rotary blood pump (CORA) extensively in animals, but only as a temporary implantable left ventricular assist device. To expand the scope of future clinical applications, we recently undertook experiments to assess the feasibility of our pump for use in a standard cardiopulmonary bypass circuit. We conducted 4 experiments in adult sheep (body weight, 40 kg): 2 with CORA and 2 with the BioMedicus pump (BP) for comparison. In all experiments, a currently used extracorporeal circuit with reservoir, filter, and membrane oxygenator (Sorin monolith) was installed, and open chest extracorporeal circulation (ECC) was performed for 6 h. Hemodynamic performance and hemolysis were evaluated. CORA provided semipulsatile systemic flow at a level comparable to that of the BP. Free plasma hemoglobin levels were slightly higher with CORA, but the decrease in platelet count was the same for both devices. There was no significant difference in the extent of blood trauma. We conclude that CORA could be successfully used for ECC with an oxygenator. Negative pressure can be prevented by our specially designed control system.

摘要

我们对无瓣搏动旋转式血泵(CORA)进行了广泛的动物实验,但仅作为临时植入式左心室辅助装置。为了扩大未来临床应用的范围,我们最近开展了实验,以评估该血泵用于标准体外循环回路的可行性。我们在成年绵羊(体重40千克)身上进行了4项实验:2项使用CORA,2项使用BioMedicus血泵(BP)作为对照。在所有实验中,均安装了当前使用的带有储液器、过滤器和膜式氧合器(索林整体式)的体外循环回路,并进行了6小时的开胸体外循环(ECC)。对血流动力学性能和溶血情况进行了评估。CORA提供的半搏动性体循环流量与BP相当。CORA组的游离血浆血红蛋白水平略高,但两种装置的血小板计数下降情况相同。血液损伤程度无显著差异。我们得出结论,CORA可成功用于配备氧合器的ECC。我们专门设计的控制系统可防止负压产生。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验