Suppr超能文献

使用乳胶珠测定精液计数室的准确性。

Accuracy of semen counting chambers as determined by the use of latex beads.

作者信息

Seaman E K, Goluboff E, BarChama N, Fisch H

机构信息

Male Reproductive Center, Department of Urology, Columbia Presbyterian Medical Center, New York, New York 10028, USA.

出版信息

Fertil Steril. 1996 Oct;66(4):662-5. doi: 10.1016/s0015-0282(16)58587-2.

Abstract

OBJECTIVE

To assess the accuracy of the Hemacytometer (Hausser Scientific, Horsham, PA), Makler (Sefi-Medical Instrument, Haifa, Israel), Cell-VU (Millennium Sciences Inc., New York, NY), and Micro-Cell chambers (Conception Technologies, San Diego, CA) counting chambers.

DESIGN

A solution containing a known concentration of latex beads was used as the standard to perform counts on the four different counting chambers.

MAIN OUTCOME MEASURES

Bead counts for the four different chambers were compared with the bead counts of the standard solution. Variability within chambers also was determined.

RESULTS

Mean bead concentrations for both the Cell-VU and Micro-Cell chambers were consistently similar to the bead concentration of the standard solution. Both the hemacytometer and the Makler chambers overestimated the actual bead concentration of the standard solution by as much as 50% and revealed significant interchamber variability.

CONCLUSIONS

Our data revealed marked differences in the accuracy and reliability of the different counting chambers tested and emphasized the need for standardization and quality control of laboratory procedures.

摘要

目的

评估血细胞计数板(豪泽科学公司,霍舍姆,宾夕法尼亚州)、马克勒计数板(塞菲医疗仪器公司,海法,以色列)、细胞观察计数板(千禧科学公司,纽约,纽约州)和微孔细胞计数板(概念技术公司,圣地亚哥,加利福尼亚州)计数室的准确性。

设计

使用含有已知浓度乳胶珠的溶液作为标准,对四种不同的计数室进行计数。

主要观察指标

将四种不同计数室的珠计数与标准溶液的珠计数进行比较。还确定了计数室内的变异性。

结果

细胞观察计数板和微孔细胞计数板的平均珠浓度始终与标准溶液的珠浓度相似。血细胞计数板和马克勒计数板均将标准溶液的实际珠浓度高估了多达50%,并显示出显著的计数室间变异性。

结论

我们的数据显示,所测试的不同计数室在准确性和可靠性方面存在显著差异,并强调了实验室程序标准化和质量控制的必要性。