Kussak A, Andersson B, Andersson K
National Institute of Occupational Health, Analytical Chemistry Division, Umeå, Sweden.
J Chromatogr B Biomed Appl. 1995 Oct 20;672(2):253-9. doi: 10.1016/0378-4347(95)00227-a.
A method for the determination of aflatoxins B1, B2, G1, G2, M1 and Q1 in human urine has been developed. The 10-ml urine samples were automatically cleaned up on immunoaffinity columns and analysed by high-performance liquid chromatography (HPLC), including post-column derivatization with bromine and fluorescence detection. Average aflatoxin recoveries were: B1 103%, B2 106%, G1 98% and G2 96% in the range 6.8-73 pg/ml of urine and M1 103% and Q1 100% in the range 18-97 pg/ml of urine. The relative standard deviations were all between 1% and 21%. The determination limits of aflatoxins in urine were 6.8 pg/ml for B1, B2, G1 and G2 and 18 pg/ml for M1 and Q1.