Bais R, Prior M P, Edwards J B
Clin Chem. 1977 Jun;23(6):1056-8.
Plasma lactate dehydrogenase activity is artifactually increased when analyzed on the SMAC (Technicon) continuous-flow analyzer. The factors responsible are incomplete centrifugation and the presence of detergent in the dilution buffer. The contribution of platelet lactate dehydrogenase in the plasma is demonstrated with a case of thrombocytosis.
当在SMAC(Technicon)连续流动分析仪上进行分析时,血浆乳酸脱氢酶活性会出现人为升高。造成这种情况的因素是离心不完全以及稀释缓冲液中存在去污剂。通过一例血小板增多症病例证明了血小板乳酸脱氢酶在血浆中的作用。