Vatnitsky S M, Siebers J V, Miller D W
Department of Radiation Medicine, Loma Linda University Medical Center, California 92354, USA.
Med Phys. 1995 Nov;22(11 Pt 1):1749-52. doi: 10.1118/1.597631.
A method has been developed for measuring photon beam quality correction factors kQ using direct transfer with a water calorimeter. kQ values were measured for beam qualities varying between 6 and 23 MV for a Capintec PR-06 and a PTW W30001 cylindrical ionization chamber. Measured values were intercompared with published sets of computed kQ values and agreement was found to be within measurement uncertainty (1%, one standard deviation).
已开发出一种使用水热计直接传递来测量光子束质量校正因子kQ的方法。针对Capintec PR - 06和PTW W30001圆柱形电离室,测量了6至23 MV之间不同光束质量的kQ值。将测量值与已发表的计算kQ值集进行相互比较,发现一致性在测量不确定度(1%,一个标准偏差)范围内。