Ruel J, Lemay J, Dumas G, Doillon C, Charara J
Biomaterials Institute, Saint-Francois d'Assise Hospital, Laval University, Québec, Canada.
ASAIO J. 1995 Oct-Dec;41(4):876-83.
The design of a new parallel plate perfusion chamber for cell behavior studies involving pulsatile flowrates is presented. It was based on fluid mechanical considerations to ensure a region of regular and uniform shear stress at the wall. A numeric solution of the flow was performed to study the effect of pulsating flow on the entrance length. Dye injection investigations in the chamber showed laminar and uniform flow in the culture region under steady state conditions.
本文介绍了一种用于涉及脉动流速的细胞行为研究的新型平行板灌注室的设计。该设计基于流体力学考虑,以确保壁面处有规则且均匀的剪切应力区域。对流动进行了数值求解,以研究脉动流对入口长度的影响。在该室中进行的染料注入研究表明,在稳态条件下培养区域内的流动呈层流且均匀。